Ionization, Static Control, Voltage Detection & Process Environment Sensor Monitoring

Ionization

Static Control

Voltage Detection & Process Environment Sensor Monitoring Instrumentation & Software

Novx Products
Frequently Asked Questions

Question:

What provisions are made to handle proximity voltage analog offset problems?

Answer:

There is an internal contact on the DART circuit board which "zeros" the proximity voltage analog signal. This offset voltage correction factor (if any) is stored in non-volatile memory. If the DART or station monitor is replaced or voltage drifts over time, entering a new correction factor is a simple matter of "re-zeroing".

Note: When the DART is connected to a Series 300, the zero jumper should remain installed (shorted).

Back to Novx Product Frequently Asked Questions

 

Need help?

Contact an Application Specialist, Sales Services, or Technical Support by sending an email to MKS, Ion Systems or call 800-367-2452 or 510-217-0600.

MKS Instruments, People's Republic of China  MKSInstruments, United Kindom  MKS Instruments, Taiwan  MKS Instruments, Germany, GmbH  MKS Instruments, Japan  MKS Instruments, France  MKS Instruments, Korea 
en español supplier support    |      MKS Logo Shop    |      privacy policy    |      legal    |      site map

© 2008 by MKS Instruments